A Primer on Electro-Static Discharge 
Teledyne API – Model T300/T300M CO Analyzer 
328 
static fields built up on other things, like you and your clothing, from discharging 
through the instrument and damaging it.   
14.4. BASIC PRINCIPLES OF STATIC CONTROL 
It is impossible to stop the creation of instantaneous static electric charges.  It is not, 
however difficult to prevent those charges from building to dangerous levels or prevent 
damage due to electro-static discharge from occurring. 
14.4.1. GENERAL RULES 
Only handle or work on all electronic assemblies at a properly set up ESD station.  
Setting up an ESD safe workstation need not be complicated.  A protective mat properly 
tied to ground and a wrist strap are all that is needed to create a basic anti-ESD 
workstation.   
 
Wrist Stra
Protective Mat
Ground Point
 
Figure 14-2: 
Basic anti-ESD Workbench 
For technicians that work in the field, special lightweight and portable anti-ESD kits are 
available from most suppliers of ESD protection gear.  These include everything needed 
to create a temporary anti-ESD work area anywhere.   
  Always wear an Anti-ESD wrist strap when working on the electronic 
assemblies of your analyzer.  An anti-ESD wrist strap keeps the person wearing it 
at or near the same potential as other grounded objects in the work area and allows 
static charges to dissipate before they can build to dangerous levels.  Anti-ESD wrist 
straps terminated with alligator clips are available for use in work areas where there 
is no available grounded plug. 
Also, anti-ESD wrist straps include a current limiting resistor (usually around one 
meg-ohm) that protects you should you accidentally short yourself to the 
instrument’s power supply.   