Teledyne t320u Manuale Utente

Pagina di 426
APPENDIX C 
Warranty/Repair Questionnaire 
T300/T300M and M300E/EM 
(04305G DCN5798)
 
 
 
TELEDYNE API CUSTOMER SERVICE 
Email: 
api-customerservice@teledyne.com
   
PHONE: (858) 657-9800     TOLL FREE: (800) 324-5190     FAX: (858) 657-9816
 
 
 
C-1 
CUSTOMER: ____________________________________  PHONE: ______________________________________  
CONTACT NAME: ________________________________  
FAX NO: ______________________________________  
SITE ADDRESS: __________________________________________________________________________________  
SERIAL NO.: ____________________________________  
FIRMWARE REVISION: __________________________  
1. Are there any failure messages? ____________________________________________________________________  
________________________________________________________________________________________________  
________________________________________________________________________________________________  
________________________________________________________________________________________________  
________________________________________________________________________________________________  
 
Please complete the following table: 
 
PARAMETER 
DISPLAYED AS 
OBSERVED 
VALUE 
UNITS 
NOMINAL RANGE 
Range 
Range 
 
PPM, MGM
1,2
  
PPB, UGM
1
 
1 – 1000 PPM
1
 
5 – 5000 PPM
2
 
Stability 
STABIL 
 
PPM 
<1.0 PPM  
with Zero Air
CO Measure 
CO MEAS 
 
mV 
2500 – 4800 MV 
CO Reference 
CO REF 
 
mV 
2500 – 4800MV 
Measure/Reference Ratio 
MR RATIO 
 
– 
1.1 – 1.3 W/ Zero Air 
Pressure 
PRES 
 
In-Hg-A 
-2”Ambient Absolute 
Sample Flow 
SAMP FL 
 
cm
3
/min 
800 ± 10% 
Sample Temp 
SAMPLE TEMP 
 
°C 
48 ± 4 
Bench Temp 
BENCH TEMP 
 
°C 
48 ± 2 
Wheel Temp 
WHEEL TEMP 
 
°C 
68 ± 2 
Box Temp 
BOX TEMP 
 
°C 
Ambient + 7 ± 10 
Photo Drive 
PHT DRIVE 
 
mV 
250 mV – 4750 mV 
Slope of CO Measurement 
CO SLOPE 
 
– 
1.0 ± .3 
Offset of CO Measurement 
CO OFFSET 
 
PPM 
0 ± 0.3 
Dark Cal Reference signal 
REF DARK OFFSET 
 
mV 
125 ± 50 mV 
Dark Cal Measurement Signal 
MEAS DARK OFFSET 
 
mV 
125 ± 50 mV 
Electric Test 
 
 
PPM 
40 ± 2 PPM 
1
 T300, M300E          
2
 T300M, M300EM           
 
06864B DCN6314