Teledyne API – Model T300/T300M CO Analyzer 
Getting Started 
61 
 
RS-485
 
C
ONNECTION 
 
As delivered from the factory, COM2 is configured for RS-232 communications.  This 
port can be reconfigured for operation as a non-isolated, half-duplex RS-485 port. Using 
COM2 for RS-485 communication disables the USB port. To configure the instrument 
for RS-485 communication, please contact the factory. 
3.3.2. PNEUMATIC CONNECTIONS 
This section provides not only pneumatic connection information, but also important 
information about the gases required for accurate calibration; it also illustrates the 
pneumatic layouts for the analyzer in its basic configuration and with options. 
Before making the pneumatic connections, carefully note the following cautionary and 
special messages: 
 
CAUTION
 
 
G
ENERAL 
S
AFETY 
H
AZARD
 
CARBON MONOXIDE (CO) IS A TOXIC GAS. 
Do not vent calibration gas and sample gas into enclosed areas. Obtain 
a Material Safety Data Sheet (MSDS) for this material.  Read and 
rigorously follow the safety guidelines described there.
 
 
 
CAUTION 
G
ENERAL 
S
AFETY 
H
AZARD
 
Sample and calibration gases should only come into contact with PTFE 
(Teflon), FEP, glass, stainless steel or brass. 
  
The exhaust from the analyzer’s internal pump MUST be vented outside 
the immediate area or shelter surrounding the instrument. 
It is important to conform to all safety requirements regarding exposure 
to CO. 