Teledyne t320u Manuale Utente

Pagina di 426
Getting Started 
Teledyne API – Model T300/T300M CO Analyzer 
74 
3.3.2.10. PNEUMATIC LAYOUT FOR ZERO SCRUBBER/ AMBIENT SPAN OPTION 
 
Figure 3-28: 
Internal Pneumatic Flow OPT 50H – Zero Scrubber/Ambient Span 
Table 3-11:  Zero/Span Valve Operating States for Option 50H 
MODE VALVE 
CONDITION 
Sample/Cal 
Open to SAMPLE inlet 
SAMPLE 
(Normal 
State) 
Zero/Span 
Open to ZERO AIR scrubber 
Sample/Cal 
Open to ZERO/SPAN valve 
ZERO CAL 
Zero/Span 
Open to ZERO AIR scrubber 
Sample/Cal 
Open to ZERO/SPAN valve 
SPAN CAL 
Zero/Span 
Open to PRESSURE SPAN inlet 
 
 
3.3.2.11. CALIBRATION GASES 
Zero air and span gas are required for accurate calibration.  
Z
ERO 
A
IR
 
Zero air is a gas that is similar in chemical composition to the earth’s atmosphere but 
scrubbed of all components that might affect the analyzer’s readings, in this case CO 
and water vapor.  If your analyzer is equipped with an Internal Zero Span (IZS) or an 
external zero air scrubber option, it is capable of creating zero air.  
06864B DCN6314