Getting Started 
Teledyne API – Model T300/T300M CO Analyzer 
80 
3.4.4.1. INTERFERENTS FOR CO MEASUREMENTS  
It should be noted that the gas filter correlation method for detecting CO is subject to 
interference from a number of other gases that absorb IR in a similar fashion to CO.  
Most notable of these are water vapor, CO
2
, N
2
O (nitrous oxide) and CH
4
 (methane).  
The T300/T300M has been successfully tested for its ability to reject interference from 
of these sources, however high concentrations of these gases can interfere with the 
instrument’s ability to make low-level CO measurements.   
For a more detailed discussion of this topic, see Section 13.2.1.3. 
3.4.4.2. INITIAL CALIBRATION PROCEDURE  
The following procedure assumes that:  
  The instrument DOES NOT have any of the available calibration valve or gas inlet 
options installed; 
  Cal gas will be supplied through the SAMPLE gas inlet on the back of the analyzer 
  The pneumatic setup matches that described in Section 3.3.2.1. 
V
ERIFYING THE 
T300/T300M
 
R
EPORTING 
R
ANGE 
S
ETTINGS
 
While it is possible to perform the following procedure with any range setting we 
recommend that you perform this initial checkout using following reporting range 
settings: 
  Unit of Measure: PPM 
  Analog Output Reporting Range: 50 ppm  
  Mode Setting: SNGL 