Teledyne API – Model T300/T300M CO Analyzer 
Getting Started 
85 
3.4.4.3.  O2 SENSOR CALIBRATION PROCEDURE 
If your T300/T300M is equipped with the optional O
2
 sensor, this sensor should be 
calibrated during installation of the instrument.  See Section 9.7.1 for instructions. 
3.4.4.4. CO2 SENSOR CALIBRATION PROCEDURE 
If your T300/T300M is equipped with the optional CO
2
 sensor, this sensor should be 
calibrated during installation of the instrument.  See Section 9.7.2 for instructions. 
Note 
Once you have completed the above set-up procedures, please fill out 
the Quality Questionnaire that was shipped with your unit and return it 
to Teledyne API.  This information is vital to our efforts in continuously 
improving our service and our products.  THANK YOU.   
 