Omega CL526 Manuale Utente

Pagina di 50
29
 Press 
the 
<2>
 key to confirm the selection and to return to the menu page
 Press 
the 
<AUTORAMP>
 key to obtain one of the following indications:
Sel : Program   
 Select 
with
 <▲> and <▼> keys the  required "Sel:program" page
 Press 
the 
<AUTORAMP> 
key to confirm the selection and to return to the menu page
 Press 
the 
<SELECT>
 key to memory load the program and to return to the operative mode
9.3.9.4
Manual step advance
To run the program with manual step advance, press the <AUTORAMP> key obtaining the following indication (as per
example indicated at par. 8.6.9.3)
.A0    -50  TcK : 
The symbol on the left of the display has the following meaning:
O =
output
 I 
=
input
 Press 
the 
<AUTORAMP>
 key to advance one step of the program. After the "end" point (n.5 in the example - B1) the
manual sequence will start again from the point 1.
  To go back to the previous step press <ENTER> + <AUTORAMP> keys
 Press 
the
 <▲> or <▼> or <SELECT> or <!
!
!
!> or <"
"
"
"> key to exit the program
9.3.10
Automatic simulation cycle
The instrument can be programmed for simulating two types of pre-programmed continuous or step ramp output.
By programming the incremental steps to its minimum value (0.1 or 1 degree resolution) the  step ramp can be
assimilated to a continuous ramp.
Select first the technical unit (°C or °F), the type of thermocouple and then follow the procedure indicated below.
The procedure will consider a simulation in mV.
9.3.10.1
Simulation cycle selection
Two different automatic simulations identified as "Autoramp1" and "Autoramp 2" can be memory stored.
To select the required program press the <SELECT> key to obtain one of the menu pages.
 Press 
the 
<AUTORAMP> 
key to obtain one of the following indications:
Sel : Autoramp 1  
Sel : Autoramp 2  
  Press the <▲> and <▼> key to select the required Autoramp program
 Press 
the 
<AUTORAMP>
 key to confirm the selection and to return to the menu page indication
 Press 
the 
<SELECT> 
key to return to the operative mode
The automatic ramp cycle has the following behaviour