Omega CL526 Manuale Utente

Pagina di 50
31
 Press 
<ENTER>+<SOAK> keys to memory store the new value
 Press 
the 
<MODE>
 key to obtain one of the following indications:
Mode 1     ^ ramp  
  single cycle program
Mode 1     ^ ramp  
  repeated cycling program
Mode 1     1 ramp  
  single cycle program
Mode +     1 ramp  
  repeated rise ramp program.
  Press the <▲> or <▼> key to select the required program type.
 Press 
<ENTER> + <MODE> keys to memory store the new selection.
  Press the <!
!
!
!> or <"
"
"
"> key to exit the set-up procedure.
9.3.10.2
Simulation cycle
  Press the <SELECT> key to obtain one of the menu pages.
  Press the <AUTORAMP> key to obtain one of the two following indications:
Sel : Autoramp 1  
Sel : Autoramp 2  
 Press 
<▲> or <▼> key to select the required program.
  Press the <AUTORAMP> to memory store the selection and to return to the menu page.
  Press the <SELECT> key to return to the normal operative mode.
  To run the automatic simulation cycle press the <AUTORAMP> key
  The display indicates the actual cycle position as shown below:
 
Prg        18.0mV  
  On the repeated cycling to stop the program press <SHIFT> + <AUTORAMP> keys.
9.3.11
Rj compensation mode check
The internal/external reference junction compensation is only enabled for temperature measurement or simulation with
thermocouples.
  During both simulation or measuring mode to check the type of reference junction mode previously installed press the
<SELECT> key to enter the type of sensor or parameter selection menu page.
  Press the <IN/OUT> key to obtain the following indication:
RJ :    22.8°C int 
The above reading indicates that the instrument is preset  with an internal automatic reference junction compensation.
The temperature indication is the value measured by the precision thin film resistance thermometer placed inside the
In/Out terminal. If, instead of the code "int" the indication