System Calibration
90003749(1) [NGA-e (CLD-Software 2.3/MLT-Software 3.1.X)] 04/98
NGA 2000
Supplement - 9
3.2 Programming Calibration Sequences
Beside the standard programs „zero calibration“ and „zero/span calibration“ it is possible
to run the system calibration in an user definable order of up to 40 steps.
to run the system calibration in an user definable order of up to 40 steps.
The menu to set-up this feature looks as follows.
SCSTEP
SCCALTYPE1
SCSTEPMOD
For programming the sequence you must
1. select „Step #“
2. select „Calibration procedure type“
3. select module
2. select „Calibration procedure type“
3. select module
Repeat this order for all program steps.
The programmable „Calibration procedure types“ are:
1. NOP
no operation (for deleting a step in an existing program)
2. ZERO-CAL
do a zero calibration
3. SPAN-CAL
do a span-calibration for all available ranges
4. ZERO/SPAN-CAL do a zero-cal then a span-cal for all available ranges
5. SPAN1-CAL
5. SPAN1-CAL
do a span calibration only for range #1
6. SPAN2-CAL
do a span calibration only for range #2
7. SPAN3-CAL
do a span calibration only for range #3
8. SPAN4-CAL
do a span calibration only for range #4
9. END-OF-PROG
end of sequence
Expert controls and set up...
↓
System Calibration...
↓
Calibration sequence programming...
↓
Calibration Sequence Programming
Step #: 1
Choose specific analyzer module...
Select all analyzer modules !
Calibration procedure type: Zero-Cal
Select all analyzer modules !
Calibration procedure type: Zero-Cal
Analyzer module:
MLT/CH3
Prg1-10
Prg11-20
Prg21-30
Prg31-40