Emerson NGA2000 Manuale Utente

Pagina di 178
System Calibration
90003749(1) [NGA-e (CLD-Software 2.3/MLT-Software 3.1.X)] 04/98
NGA 2000
Supplement - 11
3.3 Set-up general parameters
One general parameter is "Calibration Procedures in Test-Mode". It is located in the menu
"Calibration/Test procedures".
SCTESTMOD
SCCALTYPE2
SCPROGSTEP
SCCALTIME1
SCCALTIME2
With this parameter we can run all the calibration procedures either in the defined mode or
in a test mode. The test mode means that the valve switching and waiting for purge times
is done in the same manner like in the normal calibration procedure.
The only difference is that the single calibrations of the modules and the appropriate times
the modules would need for the calibration are not done.
Expert controls and set up...
System Calibration...
Calibration/Test procedures...
Calibration/Test Procedures
Start Zeroing all ranges !
Start Zeroing and Spanning all ranges !
Start calibration program !
Cancel calibration !
Calibration Procedures in Test-Mode:                       No
Test Procedures...
Calibration type:                                                 Zero-Cal
Program step:                                                                 1
Calibration time:                                                        16 s
Previous calibration time:                                         57 s
RESULTS