System Calibration
Supplement - 12
NGA 2000
90003749(1) [NGA-e (CLD-Software 2.3/MLT-Software 3.1.X)] 04/98
A further general parameter is "Timeout for Gastest". It is located in the menu "Test
procedures".
SCTESTGAS
SCTIMEOUT
SCSTEPMOD
This parameter is related to the possibility to activate a gas valve of a specific module for
test purposes. Here we can determine a time after which an activated gastest
automatically switches back onto SAMPLE-GAS.
Is this parameter set to "0" no automatic back switching is done.
Expert controls and set up...
System Calibration...
Calibration/Test Procedures...
Test procedures...
Test Procedures
Gastest of specific module:                          Sample-Gas
Timeout for gastest (0 = no timeout):                      180 s
Choose specific analyzer module...
Procedure time:                                                          61 s
Analyzer-module:
                          MLT/CH3