Agilent Technologies 60501B Manuale Utente

Pagina di 11
8
Table 60501-4.  Calibration Information
Ranges and
Calibration Points
Variables
Variables
Value
Power Supply
Settings
Current
Shunt
High Current Range
Hi_curr_rng
30
5 V/31 A
100 A
High Current Offset
Hi_curr_offset
0.013
Low Current Range
Lo_curr_rng
3
5 V/10 A
15 A
Low Current Offset
Lo_curr_offset
0.013
Voltage Range
N/A
N/A
61 V/2 A
N/A
Voltage Hi point
Volt_hipt
60
Voltage Lo point
Volt_lopt
2.7
Low Resistance Range
Lo_res_rng
2
15 V/5.5 A
15 A
Low Resistance Hi point
Lo_res_hipt
1.9
Low Resistance Lo point
Lo_res_lopt
0.067
Middle Resistance Range
Mid_res_rng
20
10.9 V/8 A
15 A
Middle Resistance Hi point
Mid_res_hipt
60
Middle Resistance Lo point
Mid_res_lopt
2.1
High Resistance Range
Hi_res_rng
2002
60 V/5 A
15 A
High Resistance Hi point
Hi_res_hipt
200
High Resistance Lo point
Hi_res_lopt
24