Xantrex Technology XDL 35-5TP Manuale Utente

Pagina di 180
  
EMC 
This instrument has been designed to meet the requirements of the EMC Directive 89/336/EEC. 
Compliance was demonstrated by meeting the test limits of the following standards: 
Emissions 
EN61326 (1998) EMC product standard for Electrical Equipment for Measurement, Control and 
Laboratory Use.  Test limits used were: 
a) 
Radiated:     Class B 
b) 
Conducted:  Class B 
c) 
Harmonics:  EN61000-3-2 (2000) Class A; the instrument is Class A by product category.  
Immunity 
EN61326 (1998) EMC product standard for Electrical Equipment for Measurement, Control and 
Laboratory Use. 
Test methods, limits and performance achieved were: 
a) 
EN61000-4-2 (1995) Electrostatic Discharge : 4kV air, 4kV contact, Performance A. 
b) 
EN61000-4-3 (1997) Electromagnetic Field, 3V/m, 80% AM at 1kHz, Performance B. 
c) 
EN61000-4-11 (1994) Voltage Interrupt, 1 cycle, 100%, Performance B.   
d) 
EN61000-4-4 (1995) Fast Transient, 1kV peak (AC line), 0·5kV peak (DC Outputs), 
Performance B. 
e) 
EN61000-4-5 (1995) Surge, 0·5kV (line to line), 1kV (line to ground), Performance B. 
f) 
EN61000-4-6 (1996) Conducted RF, 3V, 80% AM at 1kHz (AC line only; DC Output 
connections <3m not tested), Performance B. 
According to EN61326 the definitions of performance criteria are: 
Performance criterion A:  ‘During test normal performance within the specification limits.’ 
Performance criterion B:  ‘During test, temporary degradation, or loss of function or 
performance which is self-recovering’. 
Performance criterion C:  ‘During test, temporary degradation, or loss of function or 
performance which requires operator intervention or system reset occurs.’ 
Where Performance B is stated it is because DC Output regulation may deviate beyond 
Specification limits under the test conditions.  However, the possible deviations are still small and 
unlikely to be a problem in practice. 
Note that if operation in a high RF field is unavoidable it is good practice to connect the PSU to 
the target system using screened leads which have been passed (together) through an absorbing 
ferrite sleeve fitted close to the PSU terminals. 
Cautions 
To ensure continued compliance with the EMC directive observe the following precautions: 
a) 
after opening the case for any reason ensure that all signal and ground connections are 
remade correctly and that case screws are correctly refitted and tightened. 
b) 
In the event of part replacement becoming necessary, only use components of an identical 
type, see the Service Manual.