Xantrex Technology XDL 35-5TP Manuale Utente

Pagina di 180
  
Installation 
Check that the instrument operating voltage marked on the rear panel is suitable for the local 
supply. Should it be necessary to change the operating voltage, proceed as follows: 
1)  Disconnect the instrument from all voltage sources. 
2)  Remove the screws which retain the top cover and lift off the cover. 
3)  Change the connections on both transformers following the appropriate diagram below: 
BROWN
BLUE
BROWN
BLUE
BROWN
115V
230V
 
4)  Refit the cover and the secure with the same screws. 
5)  To comply with safety standard requirements the operating voltage marked on the rear panel 
must be changed to clearly show the new voltage setting. 
6)  Change the fuse to one of the correct rating, see below. 
Fuse 
The AC fuse is located in the fuse drawer in the lower part of the IEC inlet connector.  To change 
the fuse remove the line cord and open the fuse drawer with a suitable too. 
The correct mains fuse type is 20 x 5mm 250V HBC time-lag with the following rating: 
for 230V operation: 
4A (T) 250V HBC 
for 115V operation: 
8A (T) 250V HBC 
Make sure that only fuses with the required current rating and of the specified type are used for 
replacement.  The use of makeshift fuses and the short-circuiting of fuseholders are prohibited. 
Mains Lead 
When a three core mains lead with bare ends is provided it should be connected as follows:- 
Brown - Mains 
Live 
Blue - Mains 
Neutral 
Green / Yellow - 
Mains Earth 
WARNING! THIS INSTRUMENT MUST BE EARTHED 
Any interruption of the mains earth conductor inside or outside the instrument will make the 
instrument dangerous. Intentional interruption is prohibited. The protective action must not be 
negated by the use of an extension cord without a protective conductor.  
Mounting 
This instrument is suitable both for bench use and rack mounting. It is delivered with feet for 
bench mounting. The front feet include a tilt mechanism for optimal panel angle. 
A rack kit for mounting XDL Series power supplies is available from the Manufacturers or their 
overseas agents.  The rack will accommodate 1, 2 or 3 single units or a triple and single unit; a 
blanking piece is also available for unused positions in the rack. 
Ventilation 
The power supply is cooled by an intelligent multi-speed fan which vents at the rear.  Take care 
not to restrict the air inlets at the side panels or the exit at the rear.  In rack-mounted situations 
allow adequate space around the instrument and/or use a fan tray for forced cooling. 
  
 
11