Xantrex Technology XDL 35-5TP Manuale Utente

Pagina di 180
  
Connections 
Front Panel Connections 
The loads should be connected to the positive (red) and negative (black) terminals marked 
OUTPUT 1, OUTPUT 2, or AUXILIARY. 
Remote sense connections to the loads on Outputs 1 or 2, if required, are made from the 
corresponding positive (+) and negative (
−) REMOTE SENSE terminals.  Remote sense 
operation is selected from the keyboard or via a remote control interface (XDL 35-5TP only); the 
REMOTE SENSE lamp is lit when remote sense is selected.  Switching off remote sense returns 
the instrument to local sensing at the output terminals. 
The terminal marked 
 
is connected to the chassis and safety earth ground. 
Rear Panel Connections 
Auxiliary Output Terminals 
The front panel AUXILIARY OUTPUT terminals are duplicated on the rear panel with screwless 
terminals marked AUXILIARY OUTPUT. 
Main Output Terminals (XDL 35-5TP only) 
The output and sense terminals are duplicated on the rear panel screw-terminal block marked 
Output +, Output 
−, Sense + and Sense − ; these connections are paralleled with their front panel 
equivalents. 
Remote sense operation is selected from the keyboard or via a remote control interface.   When 
the rear panel terminals are used, remote sense should always be selected to ensure that output 
regulation is maintained within specification. 
RS232 (XDL 35-5TP only) 
9
−pin D−connector compatible with addressable RS232 use. The pin connections are shown 
below: 
Pin 
Name 
Description 
− 
No internal connection 
TXD 
Transmitted data from instrument 
RXD 
Received data to instrument 
− 
No internal connection 
5 GND 
Signal 
ground 
− 
No internal connection 
RXD2 
Secondary received data 
TXD2 
Secondary transmitted data 
9 GND 
Signal 
ground 
Pin 2, 3 and 5 may be used as a conventional RS232 interface with XON/XOFF handshaking. 
Pins 7, 8 and 9 are additionally used when the instrument is used in addressable RS232 mode. 
Signal grounds are connected to instrument ground. The RS232 address is set from the 
keyboard. 
 12