Alarm output 'open' for main Output off, 'closed' for main Output on 
21 
Alarm output 'closed' when Aux Output is in Current Limit.   
Set with control assigned to Output 1; applies to Output 1 alarm only. 
22 
Alarm output 'closed' when over-temperature trip occurs 
23 
Alarm output 'closed' when sense trip occurs 
24 
Alarm output 'closed' when over-current trip occurs 
25 
Alarm output 'closed' when over-voltage trip occurs 
26 
Alarm output 'closed' when any trip occurs (factory default) 
30 Buzzer 
off 
31 
Buzzer on (factory default).  A single beep indicates confirmation, a double beep 
indicates a wrong entry. 
40 
Jog digit flashes, JOG indicator  only flashes when jog is 'hidden' 
41 
JOG indicator always flashes, except when 'hidden' (factory default) 
42 
JOG indicator doesn't flash, except when 'hidden' (lazy flash) 
91 
Loads default calibration parameters.  Refer to Service Manual 
92 
Shows firmware version number in the display 
93 
Sets these # settings to their factory default 
99 
Enter calibration mode.  Refer to Service Manual. 
Factory Default Settings 
The ex-factory default settings (which will apply at first power-up) are as follows: 
Range: 35V/3A 
Voltage: 1·000V 
Current Limit:  1·000A 
OVP: 40V 
OCP: 5·5A 
Output: 
Output off; local Sense 
# Settings: 
00 
Main Output always off at power-up 
 
02 
Aux Output always off at power up (an Output 1 # function only). 
 
26 
Alarm output 'closed' when any trip condition occurs 
 31 
Buzzer 
on 
 
41 
JOG indicator always flashes; 'lazy' flash when hidden 
RS232: 
9,600 Baud (XDL 35-5TP only) 
Address: 
11 (XDL 35-5TP only) 
Error Messages 
The following hardware errors are indicated by showing the appropriate error number in the display.  
The OK key will flash and if pressed the error will be ignored and operation will continue as described. 
Error No.  Error Description 
Action on pressing OK 
Calibration constants corrupted at power-up  Loads default calibration parameters 
# functions corrupted at power-up 
Loads default # settings 
Power-down settings not correctly loaded at 
power up 
Loads factory default power-up 
settings 
Switching the instrument off with the error message showing will leave all settings unchanged. 
  
 