Emerson NGA2000 Manuale Utente
6 - 34
NGA 2000
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
Menu "Pressure Limits":
Menu "Temperature Limits":
Set up parameters:
♦
Select any line of variables with the
↓↓
-key or the
↑↑
-key.
♦
Select the variable with the ENTER key or the
→
→
-key.
♦
Select any digit with the
←
←
-key or the
→
→
-key and adjust a new value with
the
↑↑
-key or the
↓↓
-key.
♦
Confirm the new value with the ENTER key or
cancel and go back to the last value with the F2 key.
cancel and go back to the last value with the F2 key.
Options:
♦
Pressures: 0 to 3000 hPa
♦
Temperatures: 0 to 500 °C for converter; 0 to 75 °C for sensor and block
HOME
ESCAPE
INFO
CLD
F1
F2
F3
F4
F5
95.0 ppm
Pressure Limits
Sample capillary upper limit:
490.0 hPa
Ozone supply upper limit:
1050 hPa
Ozone supply lower limit:
700.0 hPa
Sample capillary lower limit:
50.0 hPa
Barometric pressure:
1013 hPa
HOME
ESCAPE
INFO
CLD
F1
F2
F3
F4
F5
95.0 ppm
Temperature limits
Converter upper limit:
500.0 C
Sensor upper limit:
50.0 C
Block upper limit:
55.0 C
Converter lower limit:
150.0 C
Block lower limit:
45.0 C
Sensor lower limit:
0.0 C