Emerson NGA2000 Manuale Utente

Pagina di 178
6 - 36
NGA 2000
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
Main Menu — Technical level configuration
↓↓
Diagnostic menus
↓↓
Analyzer module diagnostics
↓↓
Flow control parameters
↓↓
In the menu "NO/NOx flow balance" you can set up or calculate an NO
x
 to NO flow
correction factor.
Since the sample flows through different paths in the NO and NO
x
 modes of analysis,
there are differences in the flows entering the reaction chamber in the two modes. So the
CLD may produce wrong results of analysis. To compensate the different flow rate you
can set up a correction factor manually or calculate it in two different methods.
Note:
♦ 
This menu page will only be available in the software revision 2.3! The correction
function will only be valid for the CLD hardware "CLD 2" without flow balance valve!
♦ 
The NO/NO
x
 compensation of the former models will have to be done with the flow
balance valve! In that case, you will not find the menu page shown above.
Set up parameters:
♦ 
Select the line of variables or the menu line with the 
↓↓
 -key or the 
↑↑
 -key.
♦ 
Select the variable or change to the submenu with the ENTER key or the 
 -key.
♦ 
Select any digit with the 
 -key or the 
 -key and adjust a new value with
the 
↑↑
 -key or the 
↓↓
 -key.
♦ 
Confirm the new value with the ENTER key or
cancel and go back to the last value with the F2 key.
HOME
ESCAPE
INFO
CLD
F1
F2
F3
F4
F5
95.0 ppm
NO/NOx flow balance
NOx correction factor:
1.0000
Calculate factor using pressure ratio...
Calculate factor using span gas response ratio...
6.3.2.5 Diagnostic Menus
Flow control parameters