System Calibration
Supplement - 14
NGA 2000
90003749(1) [NGA-e (CLD-Software 2.3/MLT-Software 3.1.X)] 04/98
3.4.1 Control via menu system
The possibility to control via menu is located in the menu page „Calibration/Test
procedures“. There can be started and stopped any kind of system calibration respectively
test procedure.
SCTESTMOD
SCCALTYPE2
SCPROGSTEP
SCCALTIME1
SCCALTIME2
During a running SYSCAL there can be watched current information:
• 
running calibration type
• 
running program step of user defined program (other modes than user program show a
„0“!)
• 
consumed calibration time
• 
calibration time of last valid SYSCAL
Expert controls and set up...
System Calibration...
Calibration/Test procedures...
Calibration/Test Procedures
Start Zeroing all ranges !
Start Zeroing and Spanning all ranges !
Start calibration program !
Cancel calibration !
Calibration Procedures in Test-Mode:                       No
Test Procedures...
Calibration type:                                                 Zero-Cal
Program step:                                                                 1
Calibration time:                                                        16 s
Previous calibration time:                                         57 s
RESULTS