Emerson NGA2000 Manual Do Utilizador
System Calibration
Supplement - 14
NGA 2000
90003749(1) [NGA-e (CLD-Software 2.3/MLT-Software 3.1.X)] 04/98
3.4.1 Control via menu system
The possibility to control via menu is located in the menu page „Calibration/Test
procedures“. There can be started and stopped any kind of system calibration respectively
test procedure.
procedures“. There can be started and stopped any kind of system calibration respectively
test procedure.
SCTESTMOD
SCCALTYPE2
SCPROGSTEP
SCCALTIME1
SCCALTIME2
SCPROGSTEP
SCCALTIME1
SCCALTIME2
During a running SYSCAL there can be watched current information:
•
running calibration type
•
running program step of user defined program (other modes than user program show a
„0“!)
„0“!)
•
consumed calibration time
•
calibration time of last valid SYSCAL
Expert controls and set up...
↓
System Calibration...
↓
Calibration/Test procedures...
↓
Calibration/Test Procedures
Start Zeroing all ranges !
Start Zeroing and Spanning all ranges !
Start calibration program !
Cancel calibration !
Calibration Procedures in Test-Mode: No
Test Procedures...
Calibration type: Zero-Cal
Program step: 1
Calibration time: 16 s
Previous calibration time: 57 s
Start Zeroing and Spanning all ranges !
Start calibration program !
Cancel calibration !
Calibration Procedures in Test-Mode: No
Test Procedures...
Calibration type: Zero-Cal
Program step: 1
Calibration time: 16 s
Previous calibration time: 57 s
RESULTS