Emerson NGA2000 Manuale Utente

Pagina di 178
5 - 4
NGA 2000
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
Main Menu — Expert controls and set up
↓↓
Analyzer module set up
↓↓
Calibration gas list
↓↓
In the menu "Calibration Gas List" you can set up the desired zero and span gas values
for all ranges. On the first menu page you will find the parameters of range 1 and 2, on the
second page you will find the parameters of range 3 and 4. You can reach the second
menu page with the F3 key (MORE).
Via the menu line "Calibration..." you can change to the menu "Zero/span calibration",
where you can start zeroing and spanning (see next page).
Set up parameters:
♦ 
Select the line of variables you want or the menu line "Calibration..." with
the 
↑↑
 -key or the 
↓↓
 -key.
♦ 
Select the parameter or change to the menu "Zero/span calibration" with
the ENTER key or the 
 -key.
♦ 
Select any digit with the 
 -key or the 
 -key and adjust a new value with
the 
↑↑
 -key or the 
↓↓
 -key.
Options zero gas:
0.00 ppm up to the end of range value
Options span gas:
0.00 ppm up to 120% of the end of range value
You can set up the end of range values
in the menu "Range Settings" (see 5.1.6 p. 5-16).
♦ 
Confirm the new value with the ENTER key or cancel and go back to the last value
with the F2 key.
HOME
ESCAPE
MORE
INFO
CLD
F1
F2
F3
F4
F5
95.0 ppm
Calibration Gas List
Zero gas - range 1:
0.00 ppm
NOx Span gas - range 1:
2.00 ppm
Zero gas - range 2:
0.00 ppm
NO Span gas - range 2:
25.00 ppm
NOx Span gas - range 2:
25.00 ppm
Calibration...
NO Span gas - range 1:
2.00 ppm
5.1.1 Analyzer Module Set up
Calibration/Calibration Gas List