12.0 CALIBRATION 
 
 
  The CBK scales are calibrated using kilogram weights and CBKa 
scales are calibrated using either kilograms or pounds depending 
on the weighing unit displayed at the time.  
 
  To  start  the  calibration,  either  get  into  the  calibration  section 
through the Scale Settings (“
FUnC 3
”- see Section 13.3) or turn the 
scale  off  and  switch  on  again  and  then  press  [Tare]  during  the 
self-test.    Enter  code  number 
0000
  and  press  [Tare].  This  will 
take you directly to the calibration section. 
 
  The display will show "
UnLoAd
"  
 
  Remove  all  weight  from  the  pan  and  then  press  the  [Tare]  key 
when  the  scale  is  stable.  After  the  Zero  point  is  set,  the  display 
will  show  “
Ld    xx
”.  Place  the  suggested  calibration  mass  on  the 
pan.  It  is  best  to  use  a  weight  close  to  the  full  capacity  of  the 
scale.  If the mass is different from the displayed value, enter the 
value of the mass in whole numbers. The kg or the lb symbol will 
be on to show the active unit. 
 
  Press the [Tare] key when the stable indicator is on.   
 
  The scale will calibrate to the mass. When complete, it will display 
PASS
”  and  then  either  display  “
S8    CAL
”  (if  entered  the  calibration 
section through the Scale Settings as per section 13.3) or return 
to  normal  weighing  (if  entered  directly).  Remove  the  calibration 
mass. 
 
  If  an  error  message  “
FAIL  H
”  or  “
FAIL  L
”  is  shown,  re-check  the 
calibration  and  repeat.  If  the  error  cannot  be  corrected  contact 
your supplier.