Page  
8
 
 
REMOTE SCALE CONNECTION 
 
The cable for the load cell goes to a 9 pin D-subminiature plug connector with 
the following connections: 
 
Pin numbers 
Connection 
Pins 1,2 
   - Excitation (0 V) 
Pins 4,5 
+ Excitation (+5 V) 
Pin 7 
+ Signal 
Pin 8 
- Signal 
  
(The sense wires connections of a six wire load cell are not used and should  
be connected to the respective Excitation pins). 
 
REMOTE SCALE SET UP 
The remote scale should set for a realistic resolution with respect to the input 
provided  by  the  load  cell/s.    See  Section  12  for  details  of  configuring  the 
remote scale. 
 
If  a  single  2  mV  /  V  load  cell is fitted and more than 60% of the load cell is 
used for full capacity then the high output of >6 mV span makes it possible to 
set a high resolution.  
 
If  this  criterion  is  met  then  the  remote  scale  can  be  set  to  a  high  resolution 
with a maximum of 1:30,000, i.e. 300 kg x 0.01 kg. 
 
Where  more  than  one  load  cell  is  fitted  or  the  total  load  cell  capacity  is  not 
utilised  then  a  reduced  resolution  should  be  selected  in  the  remote  scale 
technical set up. For example, if a system uses four 2 mV / V 1000 kg load 
cells for a scale of 1000 kg capacity then the span output at full scale will be 
only 2.5 mV.  
 
In  this  situation  the  resolution  should  be  reduced  to  give  a  good  number  of 
ADC counts per displayed division.   ie. Set to 1:5000 or 1000 kg x 0.2 kg. 
 
Setting  a  high  resolution  without  providing  a  good  input  to  the  remote  scale 
ADC  will  not  give  better  accuracy  and  may  make  the  scale  difficult  to  meet 
performance specification. 
 
For best performance ensure a minimum of 0.
1 µ
V / d.