Adam Equipment 302666242 Manuale Utente

Pagina di 44
 
© Adam Equipment Company 2011 
Page  
16
 
 
6.3  PARTS COUNTING 
 
The basic function of parts counting is same for both the scales. In order to 
do parts counting it is necessary to know the average weight of the items to 
be  counted.  This  can  be  done  either  by  weighing  a  known  number  of  the 
items and letting the scale determine the average unit weight or by manually 
inputting a known unit weight using the keypad. 
 
Either  of  the  scales  can  be  used  to  determine  the  sample  unit  weight  or  for 
the manual entry which can be used to count using either of the scales.  
 
It  is  possible  to  increase  the  accuracy  of  the  unit  weight  at  any  time  during 
the  counting  process  by  entering  the  count  displayed  and  then pressing the 
[Smpl]  key.  You  must  ensure  that  the  quantity  displayed  matches  the 
quantity  on  the  scale  before  pressing  the  key.  The  unit  weight  can  be 
adjusted  based  on  a  larger  sample  quantity.  This  will  give  greater  accuracy 
when counting larger sample sizes. 
 
6.3.1 Weighing a sample to determine the Unit Weight 
 
To  determine  the  average  weight  of  the  items  to  be  counted  place  a known 
quantity of the items on the scale and then enter the quantity being weighed. 
The  scale  will  then  divide  the  total  weight  by  the  number  of  samples  and 
display the average unit weight. 
 
  Zero the scale by pressing the [Zero] key if necessary. If a container is 
to  be  used,  place  the  container  on  the  scale  and  tare  as  discussed 
earlier. 
 
  Place  a  known  quantity  of  items  on  the  scale.  After  the  “Weight” 
display  is  stable  enter  the  quantity  of  items  using  the  numeric  keys 
followed  by  pressing  the  [Smpl]  key.  The  number  of  units  will  be 
displayed on the "Count" display and the computed average weight will 
be shown on the "Unit Weight" display.  
 
  As  more  items  are  added  to  the  scale,  the  weight  and  the  count  will 
increase. 
 
  If the scale is not stable, the calculation will not be completed.  
 
  If  the  weight  is  below  zero,  the  “Count”  display  will  show  negative 
count.