8.0  CALIBRATION
 
The PGL series can only be calibrated with an external mass. 
8.1 
MANUAL CALIBRATION 
Pressing the [Cal] key will start calibration. Calibration can also be called for after a 
set time period as determined by the user (see section 12.5). 
Calibration using External Calibration mass: 
1)
 
Pressing [Esc] will abort the calibration at any time 
2)
 
Check the display is at zero. Tare if necessary 
3)
 
Press the [Cal] key 
4)
 
The  display  will  show  the  balance  setting  a  new  Zero  condition  by  showing 
“LOAD 0”. Make sure the pan is empty then press the [Enter] key to continue 
5)
 
The  balance  will  then  show  the  value  of  the  calibration  mass  required 
sounding a beep, for example “LOAD 1000 g”  
6)
 
If  a  different  mass  is  desired  press  the  up  or  down  arrow  key  to  show  the 
alternate masses possible for the scale.   
 
 