Place the mass on the balance. Press [Enter] to continue 
8)
 
On  the  PGL  203  and  PGL  303  models  the  display  will  show  the  busy  symbol 
and  after  calibration  is  complete  it  will  show    “LOAd  0”  on  the  display.    
Remove the weight.  The scale will return to weighing. 
9)
 
All other models will return to weighing as soon as calibration is complete. 
8.2 
AUTOMATIC CALIBRATION 
The  balance  will  ask  for  calibration  when  the  balance  has  automatic  calibration 
enabled and the conditions of the automatic calibration have been met.  
The calibration will be called for if the time since last calibration exceeds a preset 
time, typically 4 hours. 
The balance will call for calibration to be carried out by flashing the “CAL” symbol 
on the display. As soon as the balance is calibrated the symbol will be turned off.   
The  Auto  calibration  feature  can  be  enabled,  disabled  or  changed  by  the  user 
options to meet the requirements of the users. See section 12.5. 
8.3 
CALIBRATION ERRORS 
Occasionally  during  calibration  an  error  will  be  detected.  These  errors  can  be 
caused by:   
 
Unstable readings 
 
Improper calibration weights being used 
 
Large shifts of zero from the factory settings 
When  an  error  is  found  a  displayed  message  will  be  shown  and  the  calibration 
must  be  done  again.    If  the  balance  has  error  messages  more  than  once  it  is 
possible the mechanics have been damaged.