Invacare IRC5PO2 Manuale Utente

Pagina di 32
SECTION 5—OPERATING INSTRUCTIONS
17
Recommended Guidelines for Optimum 
Performance
NOTE: Ensure that your concentrator is at least three inches away 
from walls, draperies or furniture to assure sufficient air flow. Avoid 
deep pile carpets and heaters, radiators or hot air registers.
Plug in Power Cord
1. Plug in power cord to an electrical outlet.
Temperature:
50°F - 95°F (10°C - 35°C)
Electrical:
No extension cords.
Placement:
No closer than three inches from the wall, 
furniture, draperies, or similar surfaces.
Tubing and Cannula:
7 ft cannula with a maximum 50 ft of 
Crush-Proof Tubing (DO NOT pinch).
Environment:
Smoke and soot-free. No confined spaces 
(Example: No closets).
Relative Humidity:
20 to 60%
Time of Operation:
Up to 24 hours per day.
Flow Rate 
Perfecto
2
 5: From 0.5 L/min to 5 L/min 
For flow rates less than 1 L/min, use the 
Pediatric Flowmeter Accessory.
Minimum Operating Time:
30 Minutes