Invacare IRC5PO2 Manuale Utente
SECTION 5—OPERATING INSTRUCTIONS
17
Recommended Guidelines for Optimum
Performance
NOTE: Ensure that your concentrator is at least three inches away
from walls, draperies or furniture to assure sufficient air flow. Avoid
deep pile carpets and heaters, radiators or hot air registers.
from walls, draperies or furniture to assure sufficient air flow. Avoid
deep pile carpets and heaters, radiators or hot air registers.
Plug in Power Cord
1. Plug in power cord to an electrical outlet.
Temperature:
50°F - 95°F (10°C - 35°C)
Electrical:
No extension cords.
Placement:
No closer than three inches from the wall,
furniture, draperies, or similar surfaces.
furniture, draperies, or similar surfaces.
Tubing and Cannula:
7 ft cannula with a maximum 50 ft of
Crush-Proof Tubing (DO NOT pinch).
Crush-Proof Tubing (DO NOT pinch).
Environment:
Smoke and soot-free. No confined spaces
(Example: No closets).
(Example: No closets).
Relative Humidity:
20 to 60%
Time of Operation:
Up to 24 hours per day.
Flow Rate
Perfecto
2
5: From 0.5 L/min to 5 L/min
For flow rates less than 1 L/min, use the
Pediatric Flowmeter Accessory.
Pediatric Flowmeter Accessory.
Minimum Operating Time:
30 Minutes