Troubleshooting Guide
No fluid flow
Dispensing tip may be clogged.
Replace dispensing tip.
Reservoir pressure may be too low.
Check pressure setting.
Material feed fitting may be clogged.
Clean or replace.
Motor inoperative. Check cable con-
nections and reset motor by turning
power switch off then on.
Motor wires may be on wrong terminals.
Check to be sure white wire is on I/O
Pin 7 and brown wire is on Pin 8. 
Fluid does not shut off
Reservoir pressure is too high. Check
pressure setting.
Air is trapped in the fluid path. Remove
dispensing tip and purge the valve of
all air.
Fluid leaks into motor housing
Replace "U" cup seal.
Inconsistent deposits
Reservoir pressure may not be set
correctly. Check pressure setting.
Check for air in the fluid.
Fluid path may be partially obstructed.
Service fluid body. 