Insulet Corporation 017 Manuale Utente

Pagina di 190
122
10
Meter Error 3
See user manual for 
further instructions.
Incorrect test procedure. 
For example, putting blood 
on the test strip before 
inserting the test strip into 
the meter or applying 
blood before the Blood 
Drop and Test Strip symbols 
display.
Problem with the test strip.
Problem with the meter.
1. Be sure you see the Blood Drop and Test Strip symbols on the display 
screen before you apply blood or control solution.
2. Conduct a control solution test using a new test strip. If the results of 
the control solution test are within the range printed on the side of 
the test strip vial, retest using blood and a new test strip.
3. If the control solution test does not work or the error persists, call 
Customer Care.
Meter Error 4
See user manual for 
further instructions.
Problem with the test strip.
Problem with the meter.
1. Conduct a control solution test using a new test strip. If the results of 
the control solution test are within the range printed on the side of 
the test strip vial, retest using blood and a new test strip.
2. If the control solution test does not work or the error persists, call 
Customer Care.
Meter Error 6
See user manual for 
further instructions.
Blood sample was applied 
while the code number was 
being changed.
1. Retest using a new test strip. Do not press any buttons or Soft Keys 
while applying blood or control solution to the test strip.
2. If the error persists, call Customer Care.
Alert error message
Possible causes
Action