CHAPTER 11
Communication Failures
■ 
The Communication Process
When you turn on or use the Personal Diabetes Manager (PDM), 
it establishes communication with the active Pod. Usually, com-
munication occurs quickly. Occasionally, communication takes 
longer and the PDM displays the communication icon during 
that time in the upper left-hand corner of the current screen.
However, if communication does not work, the PDM will walk 
you through the necessary steps to re-establish communication.
Communication can fail if the PDM is:
Too far from the Pod—The PDM and Pod should be side by 
side while priming during activation.
Interrupted by outside interference—See the “OmniPod 
System Notice Concerning Interference”
 in the Appendix.
■ 
Failure During Pod Activation or Operation
When you send an instruction from the PDM—such as
 
Start to 
begin bolus delivery or Enable to begin using a different basal 
program—the Pod usually responds quickly. However, if the 
PDM cannot send your instruction within a few seconds, it dis-
plays the “Communication error” screen (Figure 11-1 on the next 
page).
1. Move the PDM closer to the Pod. The PDM will try to commu-
nicate with the Pod again. If communication still fails, you will 
see the message in Figure 11-2 on the next page.