Manual Shut-off of a Hazard Alarm
If you respond to a hazard alarm but the PDM cannot communi-
cate with the Pod to acknowledge or shut off the alarm, you can 
shut off the alarm manually.
1. Remove the Pod from your body if you have not already 
done so.
2. Press the end of a paper clip or similar item straight down into 
the manual alarm shut-off port (Figure 10-2) on the bottom 
of the Pod. The port can be found by peeling back the adhe-
sive pad from the bottom of the Pod at the square end. The 
alarm shut-off port is to the right of the fill port.
3. Press the paper clip straight down into the manual alarm shut 
off port until the alarm stops and can no longer be heard. 
Then discard the Pod according to local waste disposal regu-
lations.
Pod deactivation required.
Press “Ok” to deactivate.
Hazard 
Alarm
PDM
Continuous tone.
This alarm occurs after a PDM error 
occurs that requires the user to reset the 
date and time. Press OK to acknowl-
edge.
Change Pod.
Check blood glucose.
Alarm
Type
Source
Beep Pattern
Action
Figure 10-1
Manual 
alarm shut-
off port
Fill
port