Insulet Corporation 017 Manuale Utente

Pagina di 190
145
■ 
Pod Specifications
Size: 1.53" wide x 2.05" long x 0.57" high
 (3.88cm x 5.2cm x 1.44cm)
Weight (with full reservoir): 30 grams (1.05 oz)
Operating temperature range: 40° F to 104° F (4.4° C to 40° C) 
Under normal circumstances, your body temperature will keep 
the Pod well within this range.
Startup temperature: above 50° F (10° C)
Storage temperature range: 32° F to 86° F (0° C to 30° C)
Reservoir volume (deliverable): 200 units
Cannula insertion depth: 6.5mm
Waterproof rating: IPX8 (25 feet for up to 60 minutes)
Insulin concentration: U-100
Alarm type: Audible
Operating relative humidity range: 20–85%, noncondensing
Storage relative humidity range: 20–85%, noncondensing
Operating atmospheric pressure: 696 hPA to 1060 hPA
Storage atmospheric pressure: 696 hPA to 1060 hPA
Non-pyrogenic: Fluid pathway only
Type BF: Medical device (protection from electrical shock)
Flow Capability:
 Basal: Programmable by the user in 0.05U pulses 
 up to 30.0 U per hour
 Bolus:
Rate: 1.5 units per minute
Range: 0.05 to 30.0 units 
Delivery accuracy (tested per IEC 60601-2-24):
 Basal: +/- 5% at rates > 0.1 U/hr
 Bolus: +/- 5% for all set values > 0.1 units
Bolus presets
7
Carbohydrate presets
36
Suspend
30 minutes to 2 hours
Low reservoir volume 
indicator
10 to 50 units in 5-unit increments
Default is 10.0 units
Pod expiration alert
1 to 24 hours in 1-hour increments
Default is 4 hours
Auto-off alert
1 to 24 hours or Off in 1-hour 
increments
Default is Off
PDM lock
On or Off
Default is Off
History storage
5400 records/90 days
Language
English