Accuracy Test Results
The following graphs are designed to show flow accuracy of the 
Pod against given time periods.
Flow rate from startup
The following graph shows changes in the flow rate over the 
stabilization period. The measurements were made using a Pod 
with a basal rate of 0.05 U/h at high operating temperature.
Flow rate error
The following trumpet curve shows the accuracy of the delivery 
rate in relation to the observation period. The measurements 
were made using a Pod with a basal rate of 0.05 U/h at high oper-
ating temperature. The overall mean percentage flow error was 
1.40%.