Recommended Separation Distances Between Portable and Mobile RF Communications Equipment 
and the System
You can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communi-
cations equipment (transmitters) and the System, as recommended below, according to the maximum output power of the com-
munications equipment.
Rated maximum output power 
of transmitter, in watts
Separation distance according to frequency of transmitter, in meters
150 kHz to 80 MHz, d = 1.17 
P
80 MHz to 800 MHz, d = 0.35 
P 800 MHz to 2.5 GHz, d = 0.7 
P
0.01
0.12
0.035
0.070
0.1
0.37
0.11
0.22
1
1.17
0.35
0.7
10
3.70
1.11
2.21
100
11.7
3.5
7.0
For transmitters rated at a maximum output power not listed above, the recommended separation distances in meters (m) can be estimated 
using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts 
(W) according to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from struc-
tures, objects, and people.