Electromagnetic Emissions
This equipment is intended for use in the electromagnetic environment specified below. The user of this equipment should assure 
that is used in such an environment.
Emissions
Compliance according to
Electromagnetic environment
RF Emissions (CISPR 11)
Group 2
The Pod and the PDM emit low level electromagnetic energy (RF) in 
order to communicate. Although unlikely, nearby electronic equip-
ment may be affected.
CISPR B 
Emissions Classification
Class B
The System is suitable for use in all establishments including domestic 
establishments.
Electromagnetic Immunity
The System is intended for use in the electromagnetic environment specified below. You should observe these requirements in the 
use of the System.
Immunity against
IEC 60601-1-2
test level
Compliance level 
(of this device)
Electromagnetic environment
ElectroStatic Discharge, 
ESD (IEC 61000-4-2)
contact discharge:
± 6 kV
air discharge: ± 8 kV
± 8 kV
± 15 kV
If floors are covered with synthetic material, try to avoid electro-
static discharges.
Power Frequency mag-
netic fields 50/60 Hz 
(IEC 61000-4-8)
3 A/m
400 A/m
Suitable for most environments. Magnetic field strengths in 
excess of 400 A/m would be unlikely except in close proximity 