Insulet Corporation 017 Manuale Utente

Pagina di 190
14
2
a. Press New.
b. Enter a start time for the segment, then press 
Next.
c. Enter an end time for the segment, then press Next.
d. Enter the correction factor, then press Next.
e. To save the new correction factor segment, press Done.
3. To see the factors as a list, press List. To return to the graphic 
view, press Graph.
4. Press Save to accept the correction factors you have set.
Set reverse correction
1. To set reverse correction, use the Up/Down Controller 
buttons to choose On or Off, then press Next.
Enter the duration of insulin action
1. Use the Up/Down Controller buttons to enter the 
duration 
of insulin action (how long insulin lasts in your body), then 
press Next.
Select bolus increment and enter maximum bolus
1. Choose a bolus increment (how much each press of the 
Up/Down Controller buttons will increase or decrease a 
bolus amount)—either 0.05, 0.10, 0.50, or 1.00 units of insu-
lin—then press Next.
2. Choose the maximum bolus you can give at one time, then 
press Next.
Set extended bolus doses
1. To set the units for an extended bolus dose, choose % for per-
cent of the total bolus, or choose Units for units of insulin. To 
turn off the extended bolus feature, choose Off. Then press 
Next.
If you enter carbs and reverse corrections are On, the 
System will subtract insulin from the suggested meal 
bolus to compensate for a low blood glucose (below 
the BG target you set). With reverse corrections Off
the System will not subtract insulin from the sug-
gested meal bolus dose due to a low BG. Follow your 
healthcare provider’s recommendation to set 
reverse correction On or Off.
The duration-of-insulin-action setting is used to 
calculate the amount of insulin on board (IOB) from a 
previous correction bolus or meal bolus when you 
use the suggested bolus calculator.
The maximum bolus default is 10 units. Check with 
your healthcare provider before adjusting these set-