Insulet Corporation 017 Manuale Utente

Pagina di 190
12
2
6. Choose a temporary basal feature, then press Next. Choose 
% to show temporary basal rates as a percent increase or 
decrease of your current basal rate. Choose U/hr to show 
temporary basal rates in units per hour. Or choose Off to turn 
off the temporary basal feature. For details about setting tem-
porary basals, see Chapter 3, Understanding and Adjusting 
Basal Rates.
Enter blood glucose sound setting and BG goal
1. To set the sound for the blood glucose meter, choose On or 
Off, then press Next.
2. Use the Up/Down Controller buttons to enter the lower 
limit for your blood glucose (BG) goal, then press Next.
3. Enter the upper limit for your BG goal, then press Next.
4. Press Save to accept the BG goal.
Set the suggested bolus calculator
1. To set the suggested bolus calculator feature, choose On or 
Off, then press Next.
Enter target blood glucose value
1. Use the Up/Down Controller button to enter a target blood 
glucose (BG) value, then press Next.
2. Use the Up/Down Controller buttons to enter the “correct 
above” value or correction threshold (the BG value above 
which you want the PDM to suggest a correction bolus), then 
press Next.
3. To save the target BG value, press Done. To change these set-
tings, select the segment you wish to change and press Edit.
You can define up to 8 target BG time segments. To enter 
additional segments, choose [add new], then:
a. Press New
You can add up to 7 basal programs and set 24 
segments per program.
The lower and upper BG limits are only used when 
viewing BG History (see Chapter 8, Understanding 
Your Records
).
Check with your healthcare provider before adjust-
ing these settings.
If the suggested bolus calculator is set to On, perform 
the following steps. If set to Off, you will be prompted 
to “Select bolus increment” and “Enter maximum 
bolus.”
If the suggested bolus calculator is On, a correction 
bolus will be suggested only when your BG is above 
the value shown in brackets on the screen.