Insulet Corporation 017 Manuale Utente

Pagina di 190
45
5
5. After you press Confirm, you will see the message in
Figure 5-4 on the next page.
To remove the old Pod:
a. Gently lift the edges of the adhesive tape from your skin 
and remove the entire Pod (Figure 5-3).
b. Use soap and water to remove any adhesive that remains 
on the skin or, if necessary, use an adhesive remover.
c. Discard the used Pod according to local waste disposal 
regulations.
6. After removing the old Pod, press Next (Figure 5-4).
7. Press Yes to fill and activate a new Pod (Figure 5-4).
Fill a new Pod
1. Use an alcohol prep swab to clean the top of the insulin vial, 
then discard the prep swab.
2. Securely twist the fill needle onto the syringe (Figure 5-5, on 
the next page).
3. Pull outward to remove the protective cap from the needle 
(Figure 5-6, on the next page). Save the cap; you will need it 
later.
Removing the Pod slowly will help to avoid 
possible skin irritation.
Figure 5-3
Figure 5-4
Check the infusion site for signs of infection. 
See “Avoid Infusion Site Infections” later in this 
chapter.
Before filling a Pod, ensure that no other Pods are 
being activated within 24 inches of your PDM.
Before filling a Pod with insulin, ensure that the Pod 
is above 50° Fahrenheit (10° Celsius). If the Pod has 
been exposed to temperatures below 50° F (10° C), 
then allow the Pod to be brought back to room tem-
perature before filling it with insulin.