6
4. Correction factor—You can edit existing time segments and 
add segments, up to a total of 8.
a. To edit an existing correction factor time segment, choose 
the segment, then press Edit. Use the Up/Down Control-
ler
 buttons to choose a new start time, end time, and cor-
rection factor. Press Next after each entry. 
b. To add a segment, choose [add new], then press New
Use the Up/Down Controller buttons to choose a start 
time, end time, and correction factor. Press Next after 
each entry.
c. When you have completed all time segments, press 
Done, then Save.
5. Reverse correction—Choose On or Off, then press Next.
6. Insulin action (duration)—Use the Up/Down Controller 
buttons to choose a duration of insulin action, then press 
Done.
7. Press OK.
Ratios/factors/targets
When the suggested bolus calculator is turned On, you can view 
and change all the settings using the Ratios/factors/targets 
menu option. Choose Review all settings, then press Select 
(Figure 6-8a ) or choose one setting from the menu, then press 
Select (Figure 6-8b). The steps are the same whether you choose 
all settings or individual ones.
Figure 6-7
Figure 6-8a
Figure 6-8b