Insulet Corporation 017 Manuale Utente

Pagina di 190
65
6
Temp basal
Choose 
%U/hr , or Off to set the mode for temporary basal 
rates and presets, then press Select.
Extended
Choose %Units, or Off to set the mode for extended boluses, 
then press Select.
Bolus increment
Choose 0.050.100.50, or 1.00 units for bolus increments, then 
press Select.
Max bolus
Use the Up/Down Controller buttons to enter the maximum 
bolus dose you can take, then press Enter.
Max basal rate
Use the Up/Down Controller buttons to enter the maximum 
basal rate you can choose, then press Enter.
■ 
Alerts and Reminders
In addition to automatic safety alarms (see Chapter 10, Alerts and 
Alarms)
, the OmniPod System offers a number of personal set-
tings to help you manage your diabetes. These features are 
optional. You can turn them on or off at any time, except for the 
alerts. Those you can set at levels that you find convenient to 
remind you to change your Pod. 
Notifications you can use are:
BG (blood glucose) reminder: Reminds you to check your blood 
glucose. Choose from On or Off. The default setting is Off. When 
the reminder is on, the PDM asks you each time you enter a bolus 
dose whether you want to set a BG reminder, and you can then 
choose a time interval, in 30-minute increments.
Pod expiration: Reminds you when the Pod is nearing the 72-
hour expiration time. Choose a period from 1 to 24 hours before 
expiration, in 1-hour increments. The default setting is 4 hours. 
You will hear 2 sets of beeps every minute for 3 minutes. The 
notification will repeat every 15 minutes until you press OK to 
acknowledge it. 
Low reservoir: Alerts you when insulin in the Pod reaches a cer-
tain level, so you can plan ahead to change the Pod. Choose a 
level from 10 to 50 units, in 5-unit increments. The default setting 
is 10 units.
Some settings have preset defaults, but all settings 
are changeable. See the Appendix for a list of Sys-
tem specifications, including preset default settings.