Insulet Corporation 019 Manuale Utente

Pagina di 190
134
10
Manual Shut-off of a Hazard Alarm
If you respond to a hazard alarm but the PDM cannot communi-
cate with the Pod to acknowledge or shut off the alarm, you can 
shut off the alarm manually.
1. Remove the Pod from your body if you have not already 
done so.
2. Press the end of a paper clip or similar item straight down into 
the manual alarm shut-off port (Figure 10-2) on the bottom 
of the Pod. The port can be found by peeling back the adhe-
sive pad from the bottom of the Pod at the square end. The 
alarm shut-off port is to the right of the fill port.
3. Press the paper clip straight down into the manual alarm shut 
off port until the alarm stops and can no longer be heard. 
Then discard the Pod according to local waste disposal regu-
lations.
4. The Reset Button is located in the PDM’s battery compart-
ment (Figure 10-3). The Reset Button allows the user to reset 
the PDM (i.e. if the screen locks up and goes into an alarm). 
The users PDM settings are not lost.
Figure 10-2
Manual 
alarm shut-
off port
Fill
port
Figure 10-3
Reset
Button