Cleaning and disinfecting
Cleaning is to keep free from dirt, stains, or impurities; unsoiled.
Disinfecting is the process of destroying pathogenic organisms 
or rendering them inert.
Keep your PDM test strip port and USB port free of debris and 
liquids. Dirt, dust, blood, and control solution can impair the 
functionality of your PDM or damage it.
If the battery compartment becomes soiled, gently shake out 
the debris or use a dry, lint-free cloth to remove it.
You should clean and disinfect your PDM at least once a week. 
You should also clean your PDM whenever it is visibly soiled. For 
both cleaning and disinfecting the same process should be fol-
lowed. Therefore, if you are intending to complete weekly 
cleaning and disinfection you should complete the process out-
lined to the right. 
Use the guidelines below to clean or disinfect your PDM:
The PDM can be cleaned and disinfected up to twice a week over 
the course of the life of the device (4 years). 
1. Be sure to turn off your PDM before cleaning or disinfecting.
2. To clean and disinfect the outside surfaces of the PDM hous-
ing, use Dispatch® Hospital Cleaner Disinfectant Towels with 
Bleach or its formula equivalent (available online through 
major retailers).
3. Following the instructions on the label of the Dispatch® Hos-
pital Cleaner Disinfectant Towels with Bleach , wipe the out-
side surfaces of your PDM. Treated surfaces must remain wet 
for a full two (2) minutes
.
4. Let your PDM air dry completely before checking your BG 
next. We recommend you perform a control solution test to 
ensure the System functions properly. See ”Performing a 
Control Solution Test” in Chapt
er 7, Checking Your Blood Glu-
cose.
5. If you see any signs of deterioration of the PDM (i.e. cracking, 
separation, discolorations), please contact Customer Care.
Do NOT use IV Prep
™ wipes, alcohol wipes, soap, 
detergent, or solvents. Never put any liquid into the 
battery compartment. The PDM is NOT waterproof. 
Do NOT immerse it or place it in or near water. 
It is recommended that the user pre-clean the device 
prior to each disinfection cycle.
While cleaning, do NOT allow debris or liquid to get 
into the test strip port, USB port, battery compart-
ment, or buttons (see Figure A-1 on the next page).