Ensure accurate control solution results
Replace the cap on the control solution bottle immediately 
after using it.
Do not add water or any liquid to control solution.
Control solution tests are specified to be accurate only when 
performed between 43° and 111°F(6° and 44°C).
The control solution test results should fall within the range 
printed on the test strip vial label.
If control solution results are out of this range, REPEAT THE TEST. 
Out-of-range control solution results
Out-of-range test results may be caused by:
Expired or bad control solution
Expired or bad test strip
Error in performing test
Malfunction of the System
Control solution test must be performed between 43° and 
111°F (6° and 44°C).
If your control solution test results continue to fall outside the 
range printed on the test strip vial:
1. The System may not be working properly.
2. Do not use the System to test your blood glucose.
3. Call Insulet’s Customer Care (24 hours/7 days): 800-591-3455.
When calling from outside the United States: 781-457-5098.
Performing a Control Solution Test
1. Grasp the bottom of the test strip with the gold side facing up 
(Figure 7-2).
Use only OneTouch® Verio™ control solution with 
the System. Other brands of control solution may 
produce inaccurate results with this System.
Do not use control solution past the expiration 
date or you may get inaccurate results.
Results from OneTouch® Verio™ control solution 
tests do not reflect your blood glucose level.
The control solution range is a target range for control 
solution only. It is not a target range for your blood 
glucose.