The OneTouch® Verio™ Blood Glucose Test Strips
Important test strip information
Store the test strip vial in a cool dry place between 41° and 
86° F (5° and 30° C). Keep away from direct sunlight and heat. 
Exposure to temperatures and/or humidity outside the stor-
age limits may cause inaccurate readings.
The expiration date is printed on the vial.
The discard date is 6 months after you first open a test strip 
vial. When you open a new vial, write the discard date on the 
label.
Do Not use after the expiration or discard date, whichever 
comes first.
Store your test strips in the vial only. To avoid damage or 
contamination, Do Not put test strips in any other container.
Open the vial only when taking out a test strip for use.
Close the vial lid tightly right away after removing a test strip. 
Use each test strip as soon as you take it out of the vial.
Do Not use test strips from a vial that is damaged or left open 
to air.
Do Not get dirt, food, or liquids on the test strip. You may 
touch the test strip anywhere on its surface with clean, dry 
hands.
Do Not bend, cut, or alter test strips in any way.
Test strips are for single use only.
Never reuse a test strip that already has blood or control 
solution on it
.
Operating temperature of the meter and strip is 6°- 44°C.
Apply only OneTouch® Verio™ Control Solution or a blood 
sample to the test strip.
Do Not put the used test strip back in the vial after taking a 
test.
Used test strips may be considered biohazardous waste in 
your area. Be sure to follow your health care professional’s 
advice or local laws when throwing them away.
Complete information on the OneTouch® Verio™ test 
strips (such as interferences and performance) can be 
found in the test strip insert. Please read all instruc-
tions provided on the test strip insert before using the 
OmniPod System and OneTouch® Verio™ accessories.
Keep the test strip vial away from children. Test strips 
are a choking hazard. Do Not swallow test strips. The 
test strip vial may contain drying agents that are 
harmful if inhaled or swallowed and may cause skin 
or eye irritation. Do Not ingest or swallow any items.