Boston Scientific Neuromodulation Corporation PSC5210W Manuale Utente
30
Physician Implant Manual
C
AUTION
: Ensure that the lead is fully inserted before tightening the
setscrew to prevent lead damage.
Note: • If the connector plug is used in port 2-R, it is still necessary
to tighten the setscrew as described.
• The wrench is torque-limited and cannot be overtightened.
• Ensure that the lead is fully inserted before tightening the
• Ensure that the lead is fully inserted before tightening the
setscrew to prevent lead damage.
• If there appears to be an obstruction when inserting the
lead, lead extension, or connector plug, use the torque
wrench to loosen (counterclockwise) the setscrew and/or
gently rotate the lead to help advance the proximal end.
wrench to loosen (counterclockwise) the setscrew and/or
gently rotate the lead to help advance the proximal end.
3. Place the IPG in the subcutaneous pocket with “This Side Up”
facing towards the skin.
4. Coil excess lead or extension under the IPG.
Note: To confirm good connections, check impedances before
tightening the setscrew.
5. Secure the IPG in the pocket by suturing through the holes in the
connector.
6. Close and dress the wound(s).