FLIR Commercial Systems Inc. MG300 Manuale Utente

Pagina di 20
18                                                   MG300-EN V2.0   4/11 
 
Meg OHMS 
Terminal Voltage
Range 
Resolution Accuracy 
Test current
Short circuit 
current 
125V (0%~+10%) 0.125~4.000 MΩ 0.001MΩ +(2%+10)  1mA @ load 
125kΩ 
≤1mA 
4.001~40.00 MΩ 0.01MΩ +(2%+10) 
40.01~400.0 MΩ 0.1MΩ +(4%+5) 
400.1~4000 MΩ 1MΩ +(5%+5) 
250V (0%~+10%) 0.250~4.000 MΩ 0.001MΩ +(2%+10)  1mA @ load 
250kΩ 
≤1mA 
4.001~40.00 MΩ 0.01MΩ +(2%+10) 
40.01~400.0 MΩ 0.1MΩ +(3%+5) 
400.1~4000 MΩ 1MΩ +(4%+5) 
500V (0%~+10%)  0.500~4.000 MΩ 0.001MΩ +(2%+10)  1mA @ load 
500kΩ 
≤1mA 
4.001~40.00 MΩ 0.01MΩ +(2%+10) 
40.01~400.0 MΩ 0.1MΩ +(2%+5) 
400.1~4000 MΩ 1MΩ +(4%+5) 
1000V 
(0%~+10%) 
1.000~4.000 MΩ 0.001MΩ +(3%+10)  1mA @ load 
1MΩ 
≤1mA 
4.001~40.00 MΩ 0.01MΩ +(2%+10) 
40.01~400.0 MΩ 0.1MΩ +(2%+5) 
400.1~4000 MΩ 1MΩ +(4%+5) 
 
 
Note:
 Accuracy specifications consist of two elements: 
• (% reading) – This is the accuracy of the measurement circuit. 