FLIR Commercial Systems Inc. RT30 Manuale Utente

Pagina di 20
  
RT30
™ Engineering Design Specification
 
 
Page 17 of 20
                  
7.0 EXTERNAL 
CONNECTORS 
7.1 E
XTERNAL 
P
ROBE 
C
ONNECTOR
 
The External Probe Connector shall be a standard RJ45-8 unshielded connector, without indicators. 
 
 
Pin # 
Signal Name 
Direction 
Description 
1 GND 
Signal 
Ground 
EXPDET 
In 
External Probe Detect 
EXPINB 
In 
External Probe Input 
4 ICPVPPB 
In  Programming 
Vpp 
5 ICPDATB 
I/O  Programming 
Data 
6 ICPCLKB 
In  Programming 
Clock 
EXTALGB 
In 
External Analog Input
 (1)
8 GND 
Signal 
Ground 
 
Notes: 1. 
Future 
expansion. 
 
 
Figure 7 – External Probe Connector 
 
 
8.0 MISCELLANEOUS 
SPECIFICATIONS 
 
8.1 D
ROP 
T
EST
 
The RT30 shall be capable of withstanding a 1.5m drop, per ISTA #2a.  The test includes one drop onto the 
weakest corner, one drop onto each edge radiating from that corner, and one drop on each face of the meter, all 
onto a flat, rigid base such as steel or concrete. 
 
 
8.2 R
ESISTANCE TO 
C
ONTAMINATION
 
Both the TU and RU shall be protected to the IP43 standards (protected against solid objects > 1mm
3
, direct 
sprays of water ± 60° from vertical). 
 
 
 
Confidential and Proprietary, Extech Instruments