Nihon Kohden Corporation ZM-940PA Manuale Utente

Pagina di 111
44
Operator's Manual  ZM-940PA
Six Electrodes
The 6-electrode method with lead II and lead V5 is effective for monitoring myocardial ischemia.
You can improve monitoring accuracy considerably by adding lead V4 to this combination.  Va
and Vb can be at any position of the standard 12 leads V1 to V6, but V4 and V5 are most
appropriate for myocardial ischemic monitoring.
RA/R
LA/L
RL/RF
LL/F
Va/Ca
Vb/Cb
Symbol
Lead Color
Electrode Position
AHA
IEC
AHA
IEC
Left infraclavicular fossa
LA
L
Black
Yellow
Right infraclavicular fossa
RA
R
White
Red
Below lowest rib on the left anterior axillary line
LL
F
Red
Green
Right anterior axillary line at the same level as
LL/F
RL
RF
Green
Black
Fifth intercostal space on the left midclavicular
line. (V4 position of standard 12 leads)
Va
Ca
Brown-
blue
White-
brown
Left anterior axillary line at the same level as Va.
(V5 position of standard 12 leads)
Vb
Cb
Brown-
orange
White-
black