Operator's Manual  ZM-940PA
Three Electrodes
• Lead MII, which is similar to standard lead II, used when ECG measurement has priority
Symbol
Lead Color
Electrode Position
AHA
IEC
AHA
IEC
Left infraclavicular fossa
LA
L
Black
Yellow
Right infraclavicular fossa
RA
R
White
Red
Below lowest rib on the left
anterior axillary line
LL
F
Red
Green
• Lead MI, which is similar to standard
lead I
Change F/LL and L/LA of the lead
MII.
• Lead MIII, which is similar to standard lead
III.
Change R/RA and L/LA of the lead MII.
If the electrode position shown above is not available due to chest surgery, attach the electrodes
to the root of the limbs or below the clavicles for stable ECG monitoring.