Shenzhen Creative Industry Co. Ltd. PC-303 Manuale Utente

Pagina di 69
User Manual for All-in-One Health Monitor 
 
 
Page IX/69 
 
Blood pressure measurement is prohibited to those who have severe hemorrhagic tendencies or with sickle cell 
disease, as partial bleeding may be caused.   
 
An appropriate cuff should be selected according to the age and arm circumference of the patient. The cuff width 
should be 2/3 of the length of the upper arm. The inflatable part should be long enough to permit wrapping 
approximately 80% of the limb. See the table below for the dimensions:     
Cuff Model 
Arm Circumference 
Small-sized Pediatric Cuff 
10cm~19cm 
Large-sized Pediatric Cuff 
18cm~26cm 
Small-sized Adult Cuff 
25cm~35cm 
Large-sized Adult Cuff 
33cm~47cm 
Note: The device is suitable for patient with more than 3 years old, and the appropriate cuff 
should be selected according to the age and arm circumference of the patient.